MRI Safety Information

For Advanced Bionics Cochlear Implants, Dominican Republic

Testing has demonstrated that the Advanced Bionics HiRes implant family is MRI Conditional. Conditions vary by geography. A patient with the implant may be safely scanned with MRI only under very specific conditions. Scanning under different conditions may result in severe patient injury.


MRI Safety Information by Type of Implant

Type of Implant Model Number MRI Field Strength (T) Spatial Gradient Field (T/m) Max Head SAR (W/kg) Max Body SAR (W/kg)
Clarion 1.0* MMT-5100 L/R MRI is contraindicated None None None
Clarion 1.2* AB-5100 L/R MRI is contraindicated None None None
Clarion CII Bionic Ear*     AB-5100H MRI is contraindicated None None None
HiRes 90K*     CI-1400 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes 90K Advantage     CI-1500 1.5T 2.5 T/m ≤ 1 W/kg ≤ 1.7 W/kg
HiRes Ultra CI-1600

1.5T

3.47 T/m 13.90 T/m*** ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T** 6.9 T/m ≤ 2.6 W/kg ≤ 2 W/kg
HiRes
Ultra 3D
CI-1601 1.5T 20 T/m ≤ 3.2 W/kg ≤ 2 W/kg
  3.0T 20 T/m ≤ 2.6 W/kg ≤ 2 W/kg

* These devices are no longer sold in the EU or North America
** For MRI the magnet has to be removed surgically
*** With magnet removed

Select the Implant name below for further MRI information

HiRes Ultra HiRes 90K Implant Family

 

HiRes™ Ultra Cochlear Implant 


Device HiRes Ultra
Instructions For Use
HiResolution Bionic Ear System: HiRes Ultra 
1.5T magnet removed
Scanning ok under certain conditions
1.5T magnet in place with MRI Antenna Coil Cover and bandaging protocol CI-7521 Scanning ok under certain conditions
3.0T magnet removed Scanning ok under certain conditions

MRI Safety Information

Testing has demonstrated that the HiRes Ultra cochlear implant is MR Conditional. Unilateral and bilateral recipients with this device can be safely scanned in an MR system meeting the following conditions:    

1.5T with the magnet in place

  • The external sound processor and headpiece are MR Unsafe and must be removed before entering a room containing an MR scanner.
  • A bandage must be applied using the bandaging protocol below.
  • Horizontal closed bore scanners with a static magnetic field of 1.5T.
  • Maximum spatial field gradient of 2,000 Gauss/cm (20 T/m).
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of ≤ 2 W/kg at 1.5T (Normal Operating Mode) for quadrature transmit RF body coils.
  • Maximum MR system reported, head averaged SAR of ≤ 3.2 W/kg at 1.5T (Normal Operating Mode) for quadrature transmit RF head coils.
  • RMS gradient field of 30 T/s and peak gradient field of 150 T/s.
  • Under the scan conditions defined above the HiRes Ultra implant is expected to produce a maximum temperature rise of <1°C after 15 minutes of continuous 1.5T scanning In MRI testing of unilateral and bilateral recipient conditions, respectively, the image artifact caused by the device extends from the HiRes Ultra implant approximately 8 cm using a gradient echo pulse sequence and >9.5 cm using a spin echo or gradient echo pulse sequence in a 1.5T scanner with the magnet in place. These artifacts may result in a loss of diagnostic information in the implant vicinity.
  • The recommended minimum duration of time post implant surgery prior to undergoing an MRI scan is 2 to 4 weeks in order to allow any inflammation to subside.
  • An MRI scan is not recommended if the patient has a fever.
  • Please note that the MRI Antenna Coil Cover, CI7521, and bandaging supplies must be on hand at the time of the MRI procedure. Please contact Advanced Bionics technical support at technicalservices@advancedbionics.com, +1-877-454-5051, or visit www.advancedbionics.com/mri prior to the MRI procedure to request these supplies.
  • Caution: The bandaging protocol with use of the MRI Antenna Coil Cover was developed and approved to prevent magnet displacement and counteract magnet torque during a 1.5T MRI procedure, but some discomfort and pain at the implant site may still be experienced. Please consult with your physician if this is an issue.
  • Caution: If discomfort persists following an MRI, please notify your physician.
  • Caution: Failure to secure the MRI Antenna Coil Cover and internal magnet in place during MRI may result in magnet displacement or the need for surgical revision.
  • Caution: Please consult with your physician prior to MRI to determine if the benefits of MRI are worthwhile over other imaging techniques.    

1.5T with the magnet removed

  • The internal magnet must be removed. See the “Surgeon Manual for the HiRes Ultra Cochlear Implant” for removal instructions.
  • The external sound processor and headpiece are MR Unsafe and must be removed before entering a room containing an MR scanner.
  • Horizontal closed bore scanners with a static magnetic field of 1.5T.
  • Maximum spatial field gradient of 2,000 Gauss/cm (20 T/m).
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of ≤ 2 W/kg at 1.5T (Normal Operating Mode) for quadrature transmit RF body coils.
  • Maximum MR system reported, head averaged SAR of ≤ 3.2 W/kg at 1.5T (Normal Operating Mode) for quadrature transmit RF head coils
  • RMS gradient field of 30 T/s and peak gradient field of 150 T/s.
  • Under the scan conditions defined above the HiRes Ultra implant is expected to produce a maximum temperature rise of <1°C after 15 minutes of continuous 1.5T scanning.
  • In MRI testing of unilateral and bilateral recipient conditions, respectively, the image artifact caused by the device extends from the HiRes Ultra implant approximately 3 cm and 4 cm using a gradient echo pulse sequence in a 1.5T scanner with the temporary nonmagnetic plug in place. These artifacts may result in a loss of diagnostic information in the implant vicinity.
  • The recommended minimum duration of time post implant surgery prior to undergoing an MRI scan is 2 to 4 weeks in order to allow any inflammation to subside.
  • An MRI scan is not recommended if the patient has a fever.    

Magnet Removal/Replacement Using the HiRes Ultra Magnet Tool Kit, CI‑1418

These tools must be sterilized prior to use. See the ”Guide for Reprocessing HiRes Ultra Reusable Tools” provided in the kit. 


HiRes Ultra Magnet Took Kit, CI-1418    

HiRes Ultra Magnet Tool Kit contains Two (2) Magnet Insertion Tools and Two (2) Magnet Pusher Tools


3.0T with the magnet removed

  • The internal magnet must be removed. See the “Surgeon Manual for the HiRes Ultra Cochlear Implant” for removal instructions.
  • The external sound processor and headpiece are MR Unsafe and must be removed before entering a room containing an MR scanner.
  • Horizontal closed bore scanners with a static magnetic field of 3.0T
  • Maximum spatial field gradient of 2,000 Gauss/cm (20 T/m)
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of ≤ 2 W/kg at 3.0T (Normal Operating Mode) for quadrature transmit RF body coils
  • Maximum MR system reported, head averaged SAR of ≤ 3.2 W/kg at 3.0T (Normal Operating Mode) for quadrature transmit RF head coils
  • RMS gradient field of 30 T/s and peak gradient field of 150 T/s
  • Under the scan conditions defined above the HiRes Ultra implant is expected to produce a maximum temperature rise of <3°C after 15 minutes of continuous 3.0T scanning.
  • In MRI testing of unilateral and bilateral recipient conditions, respectively, the image artifact caused by the device extends from the HiRes Ultra implant approximately 5 cm using a spin echo pulse sequence in a 3.0T scanner with the temporary nonmagnetic plug in place. These artifacts may result in a loss of diagnostic information in the implant vicinity
  • The recommended minimum duration of time post implant surgery prior to undergoing an MRI scan is 2 to 4 weeks in order to allow any inflammation to subside.
  • An MRI scan is not recommended if the patient has a fever.    

Magnet Removal/Replacement Using the HiRes Ultra Magnet Tool Kit, CI‑1418

These tools must be sterilized prior to use. See the ”Guide for Reprocessing HiRes Ultra Reusable Tools” provided in the kit. 


HiRes Ultra Magnet Took Kit, CI-1418    

HiRes Ultra Magnet Tool Kit contains Two (2) Magnet Insertion Tools and Two (2) Magnet Pusher Tools



MRI Antenna Coil Cover

The MRI Antenna Coil Cover is intended to be used along with a bandaging protocol to allow the patient to undergo an MRI procedure with the magnet left in place. Instructions for use must be followed, including the recommended bandaging protocol.

Compression Dressing prior to MRI Procedure with HiRes 90K Users

This document will provide general guidance for applying appropriate compression dressing over the site of the implanted device (and magnet) in order to perform an MRI procedure.    


Required Materials


Hurt-Free Wrap – Hospital Grade

  • Latex Free
  • 5 cm width
  • Johnson and Johnson (or equivalent)


Flexible Tape

  • Latex Free
  • 2 – 2.5 cm width
  • 3M – Nexcare (or equivalent)


MRI Antenna Coil Cover CI-7521 - Advanced Bionics



Patient Headpiece



Marking Pen



Coach Self-Adhering Elastic Bandage 3 inch (8 cm) width

  • Johnson and Johnson


MRI Antenna Coil Cover CI-7521 - Advanced Bionics



Patient Headpiece



Marking Pen




Advanced Bionics HiRes Ultra cochlear implant, magnet indicated.



Lateral view of a cochlear implant looking out toward the pinna of the ear showing magnet. Magnet must be covered by MRI Antenna Coil Cover.



Lateral x-ray in sagittal plane shows a cochlear implant in typical orientation relative to a patient’s skull with implant magnet indicated. Magnet must be covered by MRI Antenna Coil Cover.


Procedure

1

Place patient in sitting position to allow access to the implant site.

NOTE: If the patient is bilaterally implanted, prepare to locate both implant devices using the following steps. Both implant locations must have compression dressing placed over their magnets.

2

Secure MRI Antenna Coil Cover over implant magnet site. Place the patient headpiece (with cable removed) over the implant site. The magnets will hold the headpiece in place.


Patient headpiece magnetically attached

3

Cut a piece of the Coach bandage that is long enough to wrap around the head once..

4

Wrap this piece around the head, so that the bandage covers the patient headpiece. Make this wrap tight.


Patient headpiece under bandage

5

Outline the position of the headpiece on the bandage using a marker or pen.


Marked bandage showing outline of headpiece

6

Slip out the patient headpiece, but keep the bandage in place.


Removing headpiece with bandage in place

7

Slide the MRI Antenna Coil Cover under the bandage, lining it up with the marked outline of the headpiece.


Inserting MRI Antenna Coil Cover 

MRI Antenna Coin Cover under marked outline

8

Measure head size and bandage length needed for compression wrapping. Take the remaining Coach bandage roll, and wrap the bandage around the head once, without stretching.


Remaining Coach bandage roll wrapped once around head

9

Mark the location on the bandage that is one full wrap around the head. This is the head circumference.


Mark bandage one full wrap

Marked head circumference around head

10

Unwrap the piece of bandage with the marked head circumference, and place it on a flat surface.


Marked head circumference

11

Unroll the remaining bandage roll. Fold over the bandage start and crease at the marked head circumference.


12

Cut or tear the remaining bandage where it overlaps with the bandage start. The resulting bandage piece length is twice the head circumference.


13

Apply compression bandaging. Wrap this cut piece, this time very tightly, by stretching the marked line an additional half turn around the head.

NOTE: After 1.5 turns, the marked line should be on the opposite side of the head from where you started the wrap.


14

Continue wrapping the bandage at 150% extension for an additional 1.5 turns, resulting in 3 full turns total.

NOTE: After 3 turns, the bandage should end where you started the wrap (further is acceptable also).



After the MRI

Once the MRI procedure has been completed, gently remove the Coach Wrap, tape, and the MRI Antenna Coil Cover.

Discard all components including the MRI Antenna Coil Cover.


HiRes™ 90k Cochlear Implant Family

Device HiRes 90k and HiRes 90K Advantage    
Instructions For Use HiResolution Bionic Ear System: HiRes 90K and 90K Advantage
1.5T magnet removed Scanning ok under certain conditions
1.5T magnet in place with MRI Antenna Coil Cover and bandaging protocol CI-7521 Scanning ok under certain conditions

Magnetic Resonance Imaging Testing

Magnetic Resonance Imaging Testing: MRI testing for the HiRes 90K Advantage cochlear implant, was performed according to two configurations:
with the internal magnet in place and with the internal magnet removed.

These tests were performed with MRI machines having a 1.5 Tesla static field, with a 64 MHz RF pulsed field, and a 0.3 Tesla static field, with a 12 MHz RF pulsed field, with the following results:

  • The diagnostic image in situ will be distorted. With worst-case scan parameters, an image shadow around the implant area was produced. The image distortion was largest in the axial scans. The MRI static field exerts a small force on the implant.
     

MRI Guidelines

  • Use only MRI systems operating at a static magnetic field strength of 0.3 Tesla or 1.5 Tesla. MRI safety at higher energy levels has not been tested.
  • Select MRI imaging parameters to ensure a specific absorption rate (SAR) less than 1.0 W/kg in the head.
  • Carefully perform continuous verbal and visual monitoring of the patient throughout the MRI procedure
     
MRI Warnings

MRI is contraindicated except under the circumstances described below. Do not allow patients with a HiRes 90K or HiRes 90K Advantage cochlear implant to be in the area of an MRI scanner unless the following conditions have been met:

  • The bandaging protocol recommended by Advanced Bionics is followed when the patient undergoes an MRI procedure with the magnet left in place or,

  • The internal magnet is surgically removed and possibly replaced with the Magnet Insert Dummy before the patient undergoes an MRI procedure.

  • The external sound processor and headpiece are removed before entering a room where an MRI scanner is located.

MRI testing performed for the HiRes 90K and HiRes 90K Advantage cochlear implant indicated that image shadowing may extend as far as 70 cm2 (with the magnet removed) and 210 cm2 (with the magnet in place) from the implant resulting in loss of diagnostic information in the implant vicinity. The extent of the shadowing may be minimized by adjusting the signal parameters.

MRI testing of the HiRes 90K and HiRes 90K Advantage cochlear implant with the internal magnet in place is only available in markets where regulatory approval has been received. Contact your Advanced Bionics representative for more information.

For additional information regarding the use of an MRI scanner with a HiRes 90K or HiRes 90K Advantage device, please contact Advanced Bionics Technical Support and please read the Magnetic Resonance Imaging (MRI) for the HiRes 90K and HiRes 90K Advantage CI (9050050269) or contact Advanced Bionics Technical Support.


Customer Service

The form has been submitted successfully!

Thank you for contacting us to assist with your hearing solutions. Count on us—a community of hearing health professionals and recipients of cochlear implants—to help you Hear and Be Heard™. A representative from Advanced Bionics will be in touch soon!

If your request needs immediate attention or you have additional questions, please contact Hear@AdvancedBionics.com.
Toll Free Phone:  +1.866.844.HEAR (4327) US and Canada

We're here to help

Our team is comprised of caring professionals and dedicated volunteers who provide you and your loved ones with information, education, and support for choosing, getting, and living with cochlear implants.

 







AB Privacy Policy AB takes your privacy seriously. We will never sell or transfer your personal information to other companies for marketing purposes.



COM-Form-4


Auditory Technical Services

Advanced Bionics provides one-on-one technical support for professionals – including surgeons, audiologists, clinicians, educators, and therapists.

D000024947 RevB